|
Compton Scatter Polarimetry
The basic physical process used to measure polarization in
the 50-300 keV energy range is Compton scattering the cross-section
for which is given by,
where,

Here no is the frequency of the incident photon, is
the frequency of the scattered photon, q
is the scattering angle of
the scattered photon measured from the direction of the incident
photon, and h is
the azimuthal angle of the
scattered photon measured from the plane containing the electric
vector of the incident photon.
For a given value of q, the scattering cross section for polarized
radiation reaches a minimum at h
= 0° and a maximum at h
= 90°. In
the case of an unpolarized beam of incident photons, there will
be no net positive electric field vector and therefore no preferred
azimuthal scattering angle (h); the distribution of scattered photons with respect to h will be uniform. However, in the polarized case, the incident photons will exhibit
a net positive electric field vector and the distribution in h will be asymmetric.
The ability to measure the polarization of the
incident photon beam depends on the asymmetry ratio.
This is defined to be the ratio between the maximum in the h distribution to the minimum intensity in the
h distribution,

This distribution is plotted
as a function of scattering angle (q)
for various incident photon energies in Figure 1. This plot shows that the asymmetry ratio is larger at lower
energies and that events with scattering angles between 60°
and 120° contain most of the polarization information.

Figure 1 - The
asymmetry ratio (eqn. 3)
for various incident photon energies.
In general, a Compton scatter polarimeter consists
of two detectors to determine the energies of both the scattered
photon and the scattered electron. One detector, the scattering
detector, provides the medium for the Compton interaction to
take place. This detector must be designed to maximize the probability
of there being a single Compton interaction with a subsequent escape
of the scattered photon. This
requires a low-Z material in order to minimize photoelectric interactions.
The area of the scattering detector which is exposed to the photon
beam is also an important factor in determining the effective area
of the polarimeter. The primary purpose of the second detector,
the calorimeter, is to aborb the full energy of the scattered photon.
The relative placement of the two
detectors defines the scattering geometry. For incident photon energies
below 100 keV, the azimuthal modulation of the scattered photons
is maximized if the two detectors are placed at a right angle relative
to the incident photon beam (q = 90°; c.f., Figure 1). The positioning
of the two detectors must also be arranged relatively close to each
other so that there is a finite solid angle for scattering to achieve
the required detection efficiency. At the same time, a larger separation
between the detectors provides more precise scattering geometry
information. The accuracy with which the scattering geometry can
be measured determines the ability to define the modulation pattern
(Figure 2) and therefore has a direct impact on the polarization
sensitivity (see below). Here, one must compromise between total
efficiency (small detector separation) and the ability to define
the modulation pattern (large detector separation).

Figure 2 - The
modulation pattern produced by
Compton scattering of polarized radiation. The
minimum of the modulation pattern defines the
plane of polarization of the incident flux.
The ultimate goal of a Compton scatter polarimeter is to measure
the azimuthal modulation pattern of the scattered photons. From equation (1), we see that the azimuthal
modulation follows a cos 2h
distribution. More
specifically, we can write the integrated azimuthal distribution
of the scattered photons as,
where j is
the polarization angle of the incident photons. (In practice, a
measured distribution must also be corrected for geometrical effects
based on the corresponding distribution for an unpolarized beam). It is customary to define, as a figure-of-merit for the polarimeter,
the polarization modulation factor. For a given energy and
incidence angle for an incoming photon beam, this can be expressed
as,

where Cp,max and Cp,min refer to the maximum and minimum number of counts registered
in the polarimeter, respectively, with respect to h; Ap and Bp refer
to the corresponding parameters in equation (4). In this case the
‘p’ subscript denotes that this refers to the measurement
of a beam with unknown polarization. In order to determine the polarization
of the measured beam, we need first to know how the polarimeter
would respond to a similar beam, but with 100% polarization. This can be done using Monte Carlo simulations.
We then define a corresponding modulation factor for an incident
beam that is 100% polarized,

Then we can then use this result, in conjunction
with the observed modulation factor (µp), to determine the
level of polarization in a measured beam,
where P is the measured
polarization. The minimum detectable polarization (MDP) can be expressed as,

where
ns is the significance
level (number of sigma), S is the total source counting rate, B
is the total background counting rate and T is the total observation
time. Improved sensitivity to source polarization can be achieved
either by increasing the modulation factor (µ100)
or by increasing the effective area of the polarimeter (thereby
increasing the source counting rate, S).
|